Apollo®

Today’s increasingly integrated healthcare environment places more demands on the anesthesia provider. In addition, the interaction between the anesthesia workplace and the rest of the hospital is increasingly complex. The Apollo anesthesia machine can simplify this complexity. With a heightened emphasis on cost reduction in healthcare, our anesthesia solutions help you reduce operating costs and maximize reimbursement.

Ease of use

Apollo can play an important role in optimizing your anesthesia workflow and reducing process costs with features such as the fully automatic self-test. Since no user interaction is required, you are free to get on with other jobs. In addition to providing you with invaluable assistance through best-in-class therapy, Apollo also offers you an open-platform architecture - enabling it to be easily integrated into your overall workflow.

Enhanced monitoring

Create your own custom-designed anesthesia workstation by adding our Infinity® Omega Solution and Innovian® Anesthesia information management system. Infinity’s Pick and Go® transport solution not only offers seamless monitoring during transport, its intelligent docking stations and ergonomic cabling system minimize transition times while helping optimize data continuity and patient care.

User friendly ergonomics

The Apollo was designed in response to feedback from anesthesiologists and CRNAs from the United States. Dräger has developed a consistent and accepted user interface between Apollo and its entire anesthesia delivery platform including the Fabius GS, Fabius® GS premium and the Fabius Tiro. Apollo has many practical and ergonomically sound features:

  • All functional elements are within easy reach of seated or standing user
  • Unique breathing bag arm design provides the ultimate in flexibility
  • Unique larger swivel writing surface
  • Integrated auxiliary O2 and pressure gauge where you need them
  • 2 deep drawers for storage (one is lockable)
  • Integrated footrest and centrally operated brake
  • Integrated and dimmable task lighting
  • Integrated power outlet for desflurane vaporizer
  • 2 or 3 vaporizer mount – Autoexclusion interlock system

Piston vs. bellows

The anesthesia ventilator has traditionally been based upon a bellows design. After many years of producing bellows-type anesthesia ventilators, Dräger adopted the piston design as the platform for the future. The following summary highlights the key features of the piston design and the rationale for adopting this technology.

Advanced ventilation

Apollo has the ventilation technology to support patients of any age and acuity. The E-Vent plus servo controlled high-speed piston ventilator works with great precision to supply a maximum peak flow far in excess of any bellows ventilator, resulting in the kind of performance previously only seen in the ICU. Additionally, unlike a bellows ventilator, it requires no drive gas. The very short response time has clear physiological benefits for your patients, while the digital precision helps you make the most of the latest ventilation modes.

Unique low-flow wizard

Apollo is designed to support the user by providing optimal low and minimal flow settings. The consistent use of low-flow anesthesia techniques can significantly reduce the cost of inhalational anesthetic agents. Reducing the fresh gas flow from 3L/min to 1L/min can result in anesthetic agent savings of up to 50%.

Automated startup

  • Simple straightforward statup makes it easier for the staff and provides complete information on device status
  • Enhanced troubleshooting with clear text directions
  • Integrated breathing system eliminates external hoses and clutter, reducing risk of misconnection and disconnection
  • Fully automatic self test lasts approximately three minutes, reflects the results to the user, and makes results available to the user at any time.

Downloads

Apollo Anesthesia Workstation Brochure, en-us
Apollo Anesthesia Workstation Brochure, en-us

Your personal anesthesia assistant

Download

Apollo Online Trainer
Apollo Online Trainer

This Training is designed to support you in learning how to operate the Apollo effectively and with confidence.

Download

Case Study: Norwalk Hospital Chooses Dräger, en-us
Case Study: Norwalk Hospital Chooses Dräger, en-us

At this Connecticut hospital, clinicians are reaping the benefits of a multi-tiered platform in one, user-friendly format.

Download

Medical Services and Training
Study: Apollo Low Flow Wizard technology
Apollo Software 4.5 Product Information, en-us
Draeger OR Solutions Brochure, en-us

Technical Data

Lung Protective Ventilation

Operational Application
Easy-to-use anesthesia machine for medical services
Patient categories
Adults, pediatrics, neonates
Fresh gas decoupling
Ventilator technology
Electronically-controlled E-Vent
APL-Technology / Precise manual ventilaton
Dräger APL-Valve
Breathing gas warmer
Standard
Low- and minimal flow setting support
Visual filling level of bag; Econometer; Low flow wizard

Ventilation Modes

Man/Spon
CPAP in Man/Spon
Volume Controlled (VC)
VC Synchronized + PS
Pressure Controlled (PC)
PC Synchronized + PS
Pressure Support
Volume Controlled AutoFlow
APRV

Features

Screen Display
12.1" TFT black/amber or colour
Vaporizers capability
up to 3
Device self test
Automatic self test
LOOPS (p/V-loop and flow/V-loop)
Yes
Smart Pilot View
Yes
Patient gas module
in/ex O2, N2O, CO2, Agent

Technical Data

Device versions
Trolley version
Breathing system
integrated; + integrated anesthetic gas monitoring
Fresh-gas mixer
Mech. Mixer with digital read-out flowtubes (manual needle valves)
Gases
O2, Air, N2O
RFID Technology
Serial interfaces
2x RS232

Safety Concept

Emergency ventilation in case of total gas supply failure
Typical battery run-time in case of power supply failure
up to 90 min
Emergency start-up procedure